Página 1 - Instructions Display Unit

KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-Mail: [email protected] Phone: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-

Página 2 - KERN KFE-TM

10 KFE-TM-BA_IA-e-1313 6 Unpacking and installation 6.1 Installation Site, Location of Use The display units are designed in a way that relia

Página 3

KFE-TM-BA_IA-e-1313 11 6.4 Transit Securing Please note, if the display unit is used together with platform with transportation lock, this tran

Página 4 - 1 Technical data

12 KFE-TM-BA_IA-e-1313 6.7 Adjustment As the acceleration value due to gravity is not the same at every location on earth, each display unit w

Página 5 - 2 Appliance overview

KFE-TM-BA_IA-e-1313 13 Call-up menu  In weighing mode press and at the same time and the first menu block will appear.  Press repe

Página 6

14 KFE-TM-BA_IA-e-1313 Procedure  Confirm menu setting by . Ensure that there are no objects on the weighing plate.  Wait for stab

Página 7

KFE-TM-BA_IA-e-1313 15 6.8 Linearization Linearity shows the greatest deviation of a weight display on the scale to the value of the respective

Página 8

16 KFE-TM-BA_IA-e-1313  Wait for stability display „STABLE“, then press . When “LoAd 3“ is displayed, put the third adjustment weight (1/3

Página 9 - 5 Transport and storage

KFE-TM-BA_IA-e-1313 17 Position of seals and jumper Access to conductor plate:  Remove seal  Open display unit  For adjustment / ac

Página 10

18 KFE-TM-BA_IA-e-1313 7 Operation 7.1 Start-up  Press and the instrument will carry out a self-test. As soon as the weight display appea

Página 11

KFE-TM-BA_IA-e-1313 19 7.5 Weighing with taring  Deposit weighing vessel. After successful stop check press the button. Zero display and ind

Página 12

2 KFE-TM-BA_IA-e-1313 GB KERN KFE-TM Version 1.3 06/2013 Operating and installation instructions Display unit Contents 1 Technical da

Página 13

20 KFE-TM-BA_IA-e-1313 1. Call up menu  In weighing mode press and at the same time and the first menu block will appear.  2. Set

Página 14

KFE-TM-BA_IA-e-1313 21 3. How to set tolerance weighing mode  Press repeatedly until is displayed.  Acknowledge with .  Press re

Página 15

22 KFE-TM-BA_IA-e-1313 7.7 Manual totalizing With this function the individual weighing values are totalized into the sum memory by pressing

Página 16

KFE-TM-BA_IA-e-1313 23 Delete weighing data:  If you see a display of zero, press and the number of weighing, followed by the total weight w

Página 17

24 KFE-TM-BA_IA-e-1313 7.8 Automatic adding-up This function is used to issue and add individual weighing values automatically to the summatio

Página 18 - 7 Operation

KFE-TM-BA_IA-e-1313 25 7.9 Animal weighing The mean value function is suitable for weighing restless loads. Menu setting:  , see chap. 11.

Página 19 - NET appear

26 KFE-TM-BA_IA-e-1313 8 Menu 8.1 Navigation in the menu Call up menu  In weighing mode press and at the same time and the first menu b

Página 20 - SET HI by pressing

KFE-TM-BA_IA-e-1313 27 Clock Clk on Display time switched on After 5 min without change of load the weight display passes to the time display.

Página 21 - BEEP is displayed

28 KFE-TM-BA_IA-e-1313 9 Service, maintenance, disposal Before any maintenance, cleaning and repair work disconnect the appliance from the op

Página 22

KFE-TM-BA_IA-e-1313 29 9.4 Error messages Error message Description Possible causes - - - - - Maximum load exceeded Unload weighing system or

Página 23

KFE-TM-BA_IA-e-1313 3 7 Operation ... 18

Página 24 - AUTO appears

30 KFE-TM-BA_IA-e-1313 10 Instant help In case of an error in the program process, briefly turn off the display unit and disconnect from power

Página 25 - ANIMAL appears

KFE-TM-BA_IA-e-1313 31 11 Installing display unit / weighing bridge Installation / configuration of the weighing system must be carried out by a

Página 26 - 8 Menu

32 KFE-TM-BA_IA-e-1313 11.3 How to connect the platform  Disconnect device from mains.  Pull load cell cable into the display unit through

Página 27

KFE-TM-BA_IA-e-1313 33 11.4 Configure display unit Call-up configuration menu:  In weighing mode press and at the same time and the first

Página 28

34 KFE-TM-BA_IA-e-1313 Configuration menu overview: Menu block Main menu Menu item Submenu Available settings / explanation Not documented

Página 29 - - - ol -

KFE-TM-BA_IA-e-1313 35 Dual range balance Confirm by , after that the following menu items are available. Position decimal point available s

Página 30 - 10 Instant help

36 KFE-TM-BA_IA-e-1313 Multi-interval balance Confirm by , after that the following menu items are available. Position decimal point avail

Página 31

KFE-TM-BA_IA-e-1313 37 KERN & Sohn GmbH D-72322 Balingen-Frommern Postfach 4052 E-Mail: [email protected] Tel: 0049-[0]7433- 9933-0 Fax: 00

Página 33

Annex page 1 of 12 EC type-approval certificate no. DK 0199.312 Descriptive annex Contents Page 1. Name and type of instrument and modules 2

Página 34

4 KFE-TM-BA_IA-e-1313 1 Technical data KERN KFE-TM Display 6-digit Solution verifiable 6.000 e Verification class III Weighing ranges 2 Divis

Página 35

Annex page 2 of 12 EC type-approval certificate no. DK 0199.312 1. Name and type of instrument and modules The weighing instrument is designate

Página 36 - Test certificate

Annex page 3 of 12 EC type-approval certificate no. DK 0199.312 The primary functions provided are detailed below. 2.2.1 Display range The weig

Página 37 - Declaration of conformity

Annex page 4 of 12 EC type-approval certificate no. DK 0199.312 2.2.9 Operator information messages The weight indicator has a number of genera

Página 38

Annex page 5 of 12 EC type-approval certificate no. DK 0199.312 3.1.1 Connecting cable between the indicator and load cell / junction box for l

Página 39 - Descriptive annex

Annex page 6 of 12 EC type-approval certificate no. DK 0199.312 4. Interfaces and peripheral equipment 4.1 Interfaces The interfaces are chara

Página 40 - 2.2 Functions

Annex page 7 of 12 EC type-approval certificate no. DK 0199.312 7. Securing and location of seals and verification marks 7.1 Securing and seal

Página 41 - Annex page 3 of 12

Annex page 8 of 12 EC type-approval certificate no. DK 0199.312 8.1.2 Inscriptions Manufacturer’s trademark and/or name and the type designatio

Página 42 - 3. Technical data

Annex page 9 of 12 EC type-approval certificate no. DK 0199.312 9. Pictures Figure 1a KFA-TM indicator without finalisation of front. Figu

Página 43 - 3.4 Documents

Annex page 10 of 12 EC type-approval certificate no. DK 0199.312 Figure 3a KFE-TM indicator without finalisation of front. Figure 3b Finali

Página 44 - 5. Approval conditions

Annex page 11 of 12 EC type-approval certificate no. DK 0199.312 Figure 5a KFC-TM indicator without finalisation of front. Figure 5b Finali

Página 45 - 8.1 Indicator

KFE-TM-BA_IA-e-1313 5 2 Appliance overview Front view: 1. Weight display 2. Keyboard Rear view: 3. Wall bracket 4. Connection power s

Página 46 - 8.1.2.1 Load receptors

Annex page 12 of 12 EC type-approval certificate no. DK 0199.312 Issued by DELTA 10. Composition of modules - illustrated

Página 48 - Annex page 10 of 12

Annex page 1 of 7 Annex to Test Certificate No. DK0199-R76-11.10 Issued by DELTA 1. Name and type of instrument The indicators KFA-TM / KFE-TM

Página 49 - Annex page 11 of 12

Annex page 2 of 7 Annex to Test Certificate No. DK0199-R76-11.10 Issued by DELTA • Weighing unstable samples • Totalisation 3. Technical data

Página 50 - Annex page 12 of 12

Annex page 3 of 7 Annex to Test Certificate No. DK0199-R76-11.10 Issued by DELTA 11), and any particular installation requirements. A load cell

Página 51

Annex page 4 of 7 Annex to Test Certificate No. DK0199-R76-11.10 Issued by DELTA 7. Tests The indicator has been tested according to EN 45501 a

Página 52 - 2.2 Function

Annex page 5 of 7 Annex to Test Certificate No. DK0199-R76-11.10 Issued by DELTA 9. Pictures Figure 1a KFA-TM indicator without finalisatio

Página 53 - 4. Interfaces

Annex page 6 of 7 Annex to Test Certificate No. DK0199-R76-11.10 Issued by DELTA Figure 3a KFE-TM indicator without finalisation of front. F

Página 54 - 5. Conditions for use

Annex page 7 of 7 Annex to Test Certificate No. DK0199-R76-11.10 Issued by DELTA Figure 5a KFC-TM indicator without finalisation of front. F

Página 55 - 8. Documentation

6 KFE-TM-BA_IA-e-1313 2.1 Keyboard overview Key Function Turn on/off Zeroing Navigation key Confirm entry Taring Navigation k

Página 56

KFE-TM-BA_IA-e-1313 7 2.1.1 Numerical input via the navigation buttons Key Function Digit selection to the left Delete Digit selection to the

Página 57 - Annex page 6 of 7

8 KFE-TM-BA_IA-e-1313 3 Basic Information (General) 3.1 Proper use The display unit acquired by you is used in combination with a weighing pl

Página 58 - Annex page 7 of 7

KFE-TM-BA_IA-e-1313 9 3.4 Monitoring of Test Resources In the framework of quality assurance the measuring-related properties of the display uni