6.2 Performance of adjustment
Ö Turn-on system with
Ö Press and at the same time, „F1Unt“ will be displayed
Ö Press repeatedly until „ProG“ is displayed
Ö Press , „Pn“ is displayed
Ö Then press , after that and subsequently , „P1rEF“ is displayed
Ö Press , „P2 CAL“ will be displayed
Ö Confirm by , „dECi“ is displayed
Ö Press repeatedly until „CAL“ is displayed
Ö Confirm by , „UnLd“ will be displayed
Ö Wait for stability display „STABLE“, then press , the value of the adjustment
weight used most recently will be displayed. (recommended adjustment weight
see chap. 1)
Ö Enter adjustment with
, and (use and to change
to the next decimal digit, use
to change the numerical value; use to
reset to „0“ the numerical value.)
Ö Press , „LoAd“ will be displayed
Ö Place the adjustment weight, after apparition of stability display „STABLE“, press
If the adjustment has been carried out successfully, the system changes over into
weighing mode, the value of the placed adjustment weight is displayed.
If an error message appears, repeat the adjustment process.
If the fault appears again, please contact your retailer.