16 KFB/KFN-TM-BA_IA-e-1220
Call up menu:
1. Switch-on balance and during the selftest press .
2. Press , , subsequently, the first menu block
„PO CHK“ will be displayed.
3. Press repeatedly until „P2 mode“ will be displayed.
For the KFB-TM model operate the adjustment switch.
4. Press and select the set weighing scales type by .
= Single-range balance
= Dual range balance
= Multi-interval balance
5. Acknowledge with .
6. Press repeatedly until „CAL“ will be displayed.
7. Confirm with and select setting „noLin“ by .