20 KFB/KFN-TM-BA_IA-e-1220
After linearisation the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will return
into weighing mode automatically.
6.10.2 Non-verified weighing systems
Call-up menu item P3 CALCalLiner, see chap. 6.9.1
Confirm by , the password query „Pn“ will be displayed.
Press , , subsequently.
Ensure that there are no objects on the weighing pan.
Wait for stability display, then press .
When “Ld 1“ is displayed, put the first adjustment weight (1/3
max) carefully in the centre of the weighing platform. Wait for
stability display, then press .
When “Ld 2“ is displayed, put the second adjustment weight
(2/3 max) carefully in the centre of the weighing platform.
Wait for stability display, then press .
When “Ld 3“ is displayed, put the third adjustment weight
(max) carefully in the centre of the weighing platform. Wait
for stability display, then press .
After a successful linearisation the balance will carry out a
self-test. Remove adjusting weight during selftest, balance
will return into weighing mode automatically.